Electrical Safety Comprehensive AnalyzerAN9638H(F)/AN9637H(F

Electrical Safety Comprehensive AnalyzerAN9638H(F)/AN9637H(F

    AN9638H(F)/AN9637H(F)/AN9635HS(F)/AN9635HS(F)/AN9616HS(F)/AN9613HS(F)

  From medical devices to appliances, AINUO AN Series all-in-one design includes the test functionality you need to ensure your product is compliant with the most rigorous international electrical safety test standards.

  The AC Hipot Test allows you to stress your product’s insulation with high voltage to ensure it will limit current flow should a fault occur.

  Output 100 mA of AC Hipot/Dielectric Withstand test current. Ainuo AN Series meets the 200 mA short circuit requirement for 500 VA Hipot Testers.

  The DC Hipot Test allows you to stress your product’s insulation and limit leakage current output by running a DC dielectric withstand test.

  The Ground Bond,Verify the integrity of your product’s protective earth ground conductor to prevent electric shock should a fault occur.

  The The Ground Continuity,Verify the presence of your product’s protective earth ground conductor to prevent electric shock should a fault occur.

  The Insulation Resistance,Determine the total resistance of your product’s insulation.

  The Leakage Current,Ensure your product’s leakage current remains at acceptable levels during 8 different fault conditions.

  The Functional Run,Verify that your product is working properly before shipment.

  ★Comprehensive: ACW/DCW/IR/GB four in one, 500VA capacity

  ★Information: Barcode recognition, programming match, data storage, query statistics;

  ★High quality: Stable operation, accurate data, quick test, simple operation;

  ★Strong expansion: Support barcode scanner, U disk storage, IN-PAD terminal

  Features

  Four in one functions: AC withstand voltage 5000V/100mA(or 40mA), DC withstand voltage 6000V/10mA, insulation resistance 2500V/9999MΩ, grounding bond resistance 64A/600mΩ(or 32A/600mΩ).

  Fast test: The setting time of four functions is 5s, the fast testing time of four functions is about 5.6s. Safety test accuracy is 2%, 9 grades arc detection.

  Automatic test: Analyzer has automatic interface and information interface, supports combined operation with peripheral equipment including barcode scanner, server, automatic test device etc., realize automatic test of safety detection.

  Intelligent management: analyzer has intelligent programmable control functions including barcode recognition, test programming match, test conditions control etc., information management functions including test data storage (local U disk), server data exchange (network port or RS232 port), data analysis (ESRS software), can meet the requirements of MES management of modern production line.

  Open expansion: Analyzer has rich expansion functions including IN-PAD IN-PAD ipad, ESRS test control software, barcode scanner and other optional components.

  Configuration information/Specificationof AN9638H(F)/AN9637H(F)

Model Basic functions Information system Shape features Remarks
AN9638H(F) ACW/DCW/IR/GB,500VA capacity U disk storage, barcode interface 3U standard casing Standard
AN9637H(F) ACW/DCW/IR/GB,200VA capacity U disk storage, barcode interface 3U standard casing Standard
AN9636HS(F) ACW/DCW/IR,500VA capacity U disk storage, barcode interface 3U standard casing Standard
AN9635HS(F) ACW/DCW/IR,200VA capacity U disk storage, barcode interface 3U standard casing Standard
AN9616HS(F) GB, 64A  U disk storage, barcode interface 3U standard casing Standard
AN9613HS(F) GB, 32A  U disk storage, barcode interface 3U standard casing Standard
Built-in3-channel high-voltage scanning AN9637H(F)/AN9638H(F)/AN9636HS(F)/AN9635H(F) is optional,can complete the test of withstand voltage and insulation of primary-ground, secondary-ground and primary-secondary Optional
External High voltage scanning of multiple channels AN9637H(F)/AN9638H(F)/AN9636HS(F)/AN9635H(F) is optional,AN960-08(F) 8-channel high-voltage scanner, multiple cascades Optional
IN-PAD intelligent terminal Intelligent test control terminal of safety test (analyzer test, data storage network transmission, barcode scanner connector) Optional

 

 













 

Features Technical indicor
AC withstand voltage test
Rated output capacity AN9637H/AN9635HS:200VA(5000V/40mA),short-circuit current more than 200mA;
AN9638H/AN9636HS:500VA(5000V/100mA),short-circuit current more than 200mA;
Output voltage range/accuracy 100~5000V,±(2%×setting +5 counts)
Output frequency range/accuracy 50Hz or 60Hz,±0.1%×setting
Alarm current setting Upper limit range/accuracy 0.10~40.00mA,±(2%×setting +5counts)(AN9638H upper limit to 100.0mA)
Lower limit range/accuracy 0.10~40.00mA,±(2%×setting +5counts)(AN9638H(F) lower limit to 100.0mA)
Time setting Range of test time 0.5~999.9s(0 is infinite)
Range of ramp-up time 0.1~999.9s
Range of ramp-down time 0.1~999.9s
Time setting accuracy ±(0.2%×setting + 1count)
Measurement meter Voltage range/accuracy 0.01~5.00kV,±(1.5%× reading +1 count)
Frequency range/accuracy 50Hz or 60Hz,±0.1%×reading
Current range/accuracy 0.10~40.00mA,±(2%×setting +5 counts)(AN9638H(F) upper limit to 100.0mA)
Time range/accuracy 0.1~999.9s,±(0.2%×reading +1 count)
Output voltage waveform/degree of distortion/ Regulation rate Sine wave,<2%(pure resistance load),<(2%×setting+5V)(from no-load to full-load)
Initial voltage setting (0%~50%)×Output voltage setting
Current compensation setting 0.000~ 40.00mA,Outomatic(AN9638H(F) upper limit to 100.0mA)
Arc detection 1~9 levels, level-9 corresponds to the  most sensitive, 0 is Off
DC withstand voltage test
Output voltage Range/accuracy 100~6000Vdc,±(2%×setting+5 counts)
Alarm current setting Upper limit range/accuracy 0~10000μA,±(2%×setting +5 counts)
Lower limit range/accuracy 0~10000μA,±(2%×setting +5 counts)
Time setting Range of test time 0.5~999.9s(0 is infinite)
Range of ramp-up time 0.4~999.9s
Range of ramp-up time 1.0~999.9s(0 is off)
Accuracy ±(0.2%×setting +1 count)
Measurement meter Voltage range/accuracy 0.01~6.00kVdc,±(1.5%×reading +1 count)
Current range/accuracy 0.0~10000µA,±(2%×reading +5 counts)
Time range/accuracy 0.1~999.9s,±(0.2%×reading +1 count)
Rated output capacity 60VA(6000Vdc/10mA)
Output voltage ripple/regulation rate <5%(pure resistance load,6000V/10mA),<(2%×setting +5V)(from no-load to full-load)
Compensating current setting 0 ~200.0μA,automatic, manual
Initial voltage setting (0%~50%)× setting Output voltage
Lower limit current for charging (0~350)μA,automatic, manual
Discharging time ≤200ms
Arc detection 1~9 levels, level-9 corresponds to the  most sensitive, 0 is Off
Insulation resistance test
Output voltage Range/accuracy 100~2500Vdc,±(2%×setting +5V)
Alarm resistance setting Range of upper/lower limit 1~9999MΩ
Accuracy 100-499V,1~1000MΩ,±(8%×setting +2 counts)
          1000~2000MΩ,±(12%×setting +2 counts)
500-2500V,1~199MΩ,±(2%×setting +2 counts),
           200~999MΩ, ± (5%×setting +2 counts)
           1000~9999MΩ, ± (15%×setting +2 counts)
Time setting Range of ramp-up time 0.1~999.9s
Delay decision time 0,0.5~999.9s(0 is infinite)
Range of slow decreasing ramp-up time 1.0~999.9s,off
Setting accuracy ±(0.2%×setting +1 count)
Measurement meter Voltage range/accuracy 100~2500Vdc,±(2%×reading +5 counts)
Resistance range/accuracy 1~9999MΩ;
100-499V,1~1000MΩ,±(8%×reading +2 counts)
          1000~2000MΩ,±(12%×reading +2 counts)
500-2500V,1~199MΩ,±(2%×reading +2 counts),
           200~999MΩ, ± (5%×reading +2 counts)
           1000~9999MΩ, ± (15%×reading +2 counts)
Time range/accuracy 0.1~999.9s,±(0.2%×reading +1 count)
Lower limit current for charging 0~3.500μA,automatic, manual
Grounding resistance test
Rated output AN9637H(F)/AN9613HS(F): Max output is 32A, max testing resistance 600mΩ,open-circuit voltage <12V。
AN9638H(F)/AN9616HS(F): Max output is 64A,max testing resistance 600mΩ,open-circuit voltage <12V。
Output current setting Range/accuracy 2.0~32.0A(or 64.0A),±(2%×setting +5 counts)
Output voltage setting Range/accuracy 3.0~10.0V ,±(2%×setting +5 counts)
Output frequency setting Range/accuracy sine,50Hz or 60Hz,±0.1%×setting
Alarm limit setting Range of resistance upper/lower limit 2.0~10.6A:0.1~600mΩ;
10.7~32.0A:0.1~R mΩ,(AN9638H(F) 64.0A, R=(6400 / set current value)mΩ, As the current range is switch, the tester will automatically calculate the maximum range.
Error ±(2%×setting +5 counts)
Compensation resistance setting Range 0~100 mΩ,automatic measurement, compensation on/off
Test time setting Range/accuracy 0.5~999.9s(0=infinite);±(0.2%×setting +1 count)
Measurement meter Current range/accuracy 2.0~32.0A(AN9638H upper limit to 64.0A),±(2%×reading+2 counts)
Voltage range/accuracy 3.0~10.0V ,±(2%×reading +5 counts)
Resistance range/accuracy 10.0~99.9~100~600mΩ;±(2%×reading +5 counts)
Time range/accuracy 0.1~999.9s;±(0.2%×setting +1 count)
Functional interface
Remote controller interface Standard, optional foot switch (start up)
Alarm lamp connector Standard, optional three-color alarm lamp ( test, pass, alarm)
Communication port Standard, network port, RS232,RS485, GPIB one of four options, defa ult RS232
Barcode interface Standard, RS232 interface, barcode automatic recognition and programming match
U disk storage connector Standard, support standard U disk connector, test data automatic st orage
Switching value interface Standard, can output the switching value of startup, stop, test, pass and alarm
Printer connector Standard, support needle printer an d micro-printer
Rear panel test output interface Standard, rear panel output interface of high-voltage and grounding tests
Built-in three channels High voltage scanner connector Optional, can complete the withstand voltage/insulation tests of primary-ground, secondary-ground, primary-secondary withstand voltage/insulation tests of primary-ground, secondary-ground, primary-secondar y
External high voltage scanner connector Optional, AN960-08 8-channel high voltage scanner, multiple cascades
General specifications
Verification and calibration environment 23℃±5℃,(45~75)%RH,little dust
Operating environment 0 ºC ~40ºC,,(5~90)%RH,little dust
Storage environment -10℃~55℃,(5~95)%RH,little dust
Power supply of system 220V±10%,50Hz±5%
Power consumption of system Max 550W
Weight About 30kg
Dimension (mm, without external package) 426(W)×133(H)×550(D)
 
Hot selling products
AC Power Supply ANFS Series
DC Power Supply AN50 series
Electrical Safety Comprehensive TesterAN9
Grid Simulation Power Supply ANGS Series